For the Klystron based microwave bench. Klystron power supply generates required beam and repeller voltage for the X Band klystron tube like 2K25. It is very stable and contains the short circuit protection circuit. Also it has amplitude and frequency modulation circuits for the generation of 1KHz square wave and the saw tooth wave.
Voltage Range : 200-450 V Continuously Variable
Current : 50 mA Max.
Regulation : Better than 0.5% for 10% variation in the main supply voltage.
Ripple : Less than 5mV rms
Repler SupplyVoltage Range : 10 V to 270V Continuously Variable with respect to Klystron Cathode.
Regulation : 0.25% for 10% variation in the main supply voltage.
Heater Supply : 6.3 V DC (Regulated)
ModulationSquare Wave : Max. Amplitude: 110V Peak to Peak, Freq. : 500Hz-2000Hz, Amplitude and frequency constantly variable
Sawtooth : Amplitude : 60 V Max. peak to peak, Freq. : 50 Hz 150 Hz, Amplitude and frequency constantly variable
Operating Voltage : 230 V 10%, 50 Hz, A.C.